Accreditation scope for

CAL 009

Trescal Norway AS

Postboks 124
4098 TANANGER

Telephone: Tlf. 51 71 95 00 Faks. 51 71 95 01
E-mail: contact.norway@trescal.com
Internet: http://www.trescal.no
The laboratory meets the requirements in

NS-EN ISO/IEC 17025:2017




Accreditation includes:

Calibration and Measurement Capabilities (CMC)

The administrative / geographical unit:
Trescal Norway AS avd Oslo
Ensjøveien 18
0661 Oslo



Permanent facility K01 Electrical quantities

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,040 %

Remarks/ref. to internal procedure: BL08.07.03 Based on "Statnetts krav til måling av sentralnettsutveksling v.1.1(2016)" and includes evaluation of metering point. cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

DC power, calibration of meter

Range
Uncertainty (k=2)
Range
1 to 100 mW
Uncertainty (k=2)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (k=2)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (k=2)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (k=2)
Range
0,19 to 1,1 nF
Uncertainty (k=2)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (k=2)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (k=2)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (k=2)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (k=2)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (k=2)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (k=2)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (k=2)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (k=2)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (k=2)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (k=2)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (k=2)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (k=2)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (k=2)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurement

Range
Uncertainty (k=2)
Range
1 to 100 mohm
Uncertainty (k=2)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (k=2)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (k=2)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (k=2)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (k=2)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (k=2)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (k=2)
Range
0 ohm
Uncertainty (k=2)
39 µohm
Range
0 to 1 ohm
Uncertainty (k=2)
85 µohm
Range
1 to 1,9 ohm
Uncertainty (k=2)
160 µohm
Range
1,9 to 10 ohm
Uncertainty (k=2)
250 µohm
Range
10 to 25ohm
Uncertainty (k=2)
74 µohm
Range
25 to 19ohm
Uncertainty (k=2)
450 µohm
Range
19 to 100ohm
Uncertainty (k=2)
180 µohm
Range
100 to 190ohm
Uncertainty (k=2)
2,9 mohm
Range
190 to 1kohm
Uncertainty (k=2)
16 mohm
Range
1 to 1,9kohm
Uncertainty (k=2)
30 mohm
Range
1,9 to 10kohm
Uncertainty (k=2)
160 mohm
Range
10 to 19kohm
Uncertainty (k=2)
300 mohm
Range
19 to 100kohm
Uncertainty (k=2)
1,6 ohm
Range
100 to 1Mohm
Uncertainty (k=2)
3,0 ohm
Range
1 to 1,9Mohm
Uncertainty (k=2)
35 ohm
Range
1,9 to 10Mohm
Uncertainty (k=2)
350 ohm
Range
10 to 19Mohm
Uncertainty (k=2)
810 ohm
Range
19 to 100Mohm
Uncertainty (k=2)
10 kohm
Range
0,1 to 1,1Gohm
Uncertainty (k=2)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 2,2 Mv
Uncertainty (k=2)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (k=2)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (k=2)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (k=2)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (k=2)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (k=2)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (k=2)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current

Range
Uncertainty (k=2)
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,9 - 4,7 µA
Range
0,12 to 1,0 A
Uncertainty (k=2)
0,023 - 0,12mA
Range
Uncertainty (k=2)
Range
Uncertainty (k=2)
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage

Range
Uncertainty (k=2)
Range
0 to 120 mV
Uncertainty (k=2)
1,0 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance

Range
Uncertainty (k=2)
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm

Remarks/ref. to internal procedure: BL08.06.03

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,020 %

Remarks/ref. to internal procedure: BL08.07.04 cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

DC Resistance, calibration of source

Range
Uncertainty (k=2)
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
BL08.06.03
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.06.03

DC Voltage, calibration of source

Range
Uncertainty (k=2)
Range
0 to 120 mV
Uncertainty (k=2)
1 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 mV
Uncertainty (k=2)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (k=2)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (k=2)
17 - 71 µV
Range
11 to 22 V
Uncertainty (k=2)
74 - 140 µV
Range
22 to 220 V
Uncertainty (k=2)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (k=2)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (k=2)
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,90 - 4,7 µA
Range
0,12 to 1 A
Uncertainty (k=2)
0,023 - 0,12 mA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (k=2)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (k=2)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (k=2)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (k=2)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (k=2)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (k=2)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (k=2)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Voltage ratio

Range
Uncertainty (k=2)
Range
HF Powermeter DC verification
Uncertainty (k=2)
Range
0 to 100 % scale
Uncertainty (k=2)
0,06 - 0,6 % point
Zero carryover (0% scale) and instrumentation (100 % scale)

Remarks/ref. to internal procedure: MP 220

AC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (k=2)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (k=2)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (k=2)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (k=2)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (k=2)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (k=2)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (k=2)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (k=2)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (k=2)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,10 %

Remarks/ref. to internal procedure: BL08.07.01 cosφ = -1 til 1 f = 45 Hz til 70Hz I = 0,01 A til 120 A U = 5 V til 520 V

Power, using power sensor

Range
Uncertainty (k=2)
Range
0,9 to 1,1 mW
Uncertainty (k=2)
3,1 µW
49 - 51 MHz in 50 ohm system and N connector.
Range
+20 to -70 dBm
Uncertainty (k=2)
0,01 - 0,69 dB
100 kHz to 18 GHz in 50 ohm system and N connector
Range
+20 to -30 dBm
Uncertainty (k=2)
0,01 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector

Remarks/ref. to internal procedure: MP 210 Measured using Power Sensor

Attenuation

Range
Uncertainty (k=2)
Range
0 to 50 dB
Uncertainty (k=2)
0,02 - 0,46 dB
100 kHz to 18 GHz in 50 ohm system with N connector

Remarks/ref. to internal procedure: MP 212 RF substitution

Cal Factor for power sensor with N-connector

Range
Uncertainty (k=2)
Range
80 to 110 %
Uncertainty (k=2)
0,3 - 0,8 % points
100 kHz to 18 GHz in 50 ohm system and N connector
Range
80 to 110 %
Uncertainty (k=2)
Range
Uncertainty (k=2)
0,4 - 1,8 % points
10 MHz to 26,5 GHz in 50 ohm system and 3,5 mm connector

Remarks/ref. to internal procedure: MP 213

Power, generated level using power splitter

Range
Uncertainty (k=2)
Range
1 mW (0 dBm)
Uncertainty (k=2)
3,1 µW
50 MHz in 50 ohm system with N connector
Range
+20 to - 70 dBm
Uncertainty (k=2)
0,04 - 0,68 dB
100 kHz to 18 GHz in 50 ohm system and N connector. The uncertainty will be greater if adaptors must be used.
Range
+10 to - 30 dBm
Uncertainty (k=2)
0,02 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector.

Remarks/ref. to internal procedure: MP 211

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,10%

Remarks/ref. to internal procedure: BL08.07.01 cosφ = -1 til 1 f = 45 Hz til 70Hz I = 0,01 A til 120 A U = 5 V til 520 V

Termocouple (TC) Source, cal of meter

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Measure, cal of source

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (k=2)
Range
-200 °C to 630 °C
Uncertainty (k=2)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01
Permanent facility K03 Time and frequency

Periode, periode time

Range
Uncertainty (k=2)
Range
0,1µs to 100ms
Uncertainty (k=2)
1,0E-9 * Ptime+trigger error

Remarks/ref. to internal procedure: BL21.04.05

Frequency, as source

Range
Uncertainty (k=2)
Range
1,5 and 10MHz
Uncertainty (k=2)
1,0E-9 * f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, as measured quantity

Range
Uncertainty (k=2)
Range
0,5Hz to 5GHz
Uncertainty (k=2)
1,0E-9*f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, calibration of meter

Range
Uncertainty (k=2)
Range
0,010Hz to 2,0MHz
Uncertainty (k=2)
5,0 µHz + 1,7 ppm

Remarks/ref. to internal procedure: BL08.04.01

Frequency, calibration of source

Range
Uncertainty (k=2)
Range
3Hz to 1 MHz
Uncertainty (k=2)
50 µHz + 50 ppm

Remarks/ref. to internal procedure: BL08.05.01

Time interval

Range
Uncertainty (k=2)
Range
1 to 200 hour
Uncertainty (k=2)
0,80s

Remarks/ref. to internal procedure: BL21.04.04

Number, number of pulses

Range
Uncertainty (k=2)
Range
1 to 10G
Uncertainty (k=2)
0,60

Remarks/ref. to internal procedure: BL21.04.06
The administrative / geographical unit:
Trescal Norway AS avd Oslo, field calibration
Ensjøveien 18
0661 Oslo



Field testing laboratory K01 Electrical quantities

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,20 %

Remarks/ref. to internal procedure: BL08.07.02 cosφ = -1 til 1 f = 45-70Hz I = 0,01-100A U = 5-520V

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,040 %

Remarks/ref. to internal procedure: BL08.07.03 Based on "Statnetts krav til måling av sentralnettsutveksling v.1.1(2016)" and includes evaluation of metering point. cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (k=2)
Range
See remark
Uncertainty (k=2)
0,020 %

Remarks/ref. to internal procedure: BL08.07.04 cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V
The administrative / geographical unit:
Trescal Norway AS avd Tananger
NorSea Base
Bygg 53
4056 Tananger



Permanent facility K01 Electrical quantities

DC power, calibration of meter

Range
Uncertainty (k=2)
Range
1 to 100 mW
Uncertainty (k=2)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (k=2)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (k=2)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (k=2)
Range
0,19 to 1,1 nF
Uncertainty (k=2)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (k=2)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (k=2)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (k=2)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (k=2)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (k=2)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (k=2)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (k=2)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (k=2)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (k=2)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (k=2)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (k=2)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (k=2)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (k=2)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurement

Range
Uncertainty (k=2)
Range
1 to 100 mohm
Uncertainty (k=2)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (k=2)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (k=2)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (k=2)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (k=2)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (k=2)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (k=2)
Range
0 ohm
Uncertainty (k=2)
39 µohm
Range
0 - 1 ohm
Uncertainty (k=2)
85 µohm
Range
1 - 1,9 ohm
Uncertainty (k=2)
160 µohm
Range
1,9 - 10 ohm
Uncertainty (k=2)
250 µohm
Range
10 - 25ohm
Uncertainty (k=2)
74 µohm
Range
25 - 19ohm
Uncertainty (k=2)
450 µohm
Range
19 - 100ohm
Uncertainty (k=2)
180 µohm
Range
100 - 190ohm
Uncertainty (k=2)
2,9 mohm
Range
190 - 1kohm
Uncertainty (k=2)
16 mohm
Range
1 - 1,9kohm
Uncertainty (k=2)
30 mohm
Range
1,9 - 10kohm
Uncertainty (k=2)
160 mohm
Range
10 - 19kohm
Uncertainty (k=2)
300 mohm
Range
19 - 100kohm
Uncertainty (k=2)
1,6 ohm
Range
100 - 1Mohm
Uncertainty (k=2)
3,0 ohm
Range
1 - 1,9Mohm
Uncertainty (k=2)
35 ohm
Range
1,9 - 10Mohm
Uncertainty (k=2)
350 ohm
Range
10 - 19Mohm
Uncertainty (k=2)
810 ohm
Range
19 - 100Mohm
Uncertainty (k=2)
10 kohm
Range
0,1 - 1,1Gohm
Uncertainty (k=2)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 2,2 mV
Uncertainty (k=2)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (k=2)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (k=2)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (k=2)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (k=2)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (k=2)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (k=2)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Voltage, calibration of source

Range
Uncertainty (k=2)
Range
0 to 120 mV
Uncertainty (k=2)
1 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance, calibration of source

Range
Uncertainty (k=2)
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
BL08.06.03
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm

Remarks/ref. to internal procedure: BL08.06.03

DC Current

Range
Uncertainty (k=2)
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,9 - 4,7 µA
Range
0,12 to 1,0 A
Uncertainty (k=2)
0,023 - 0,12mA
Range
Uncertainty (k=2)
Range
Uncertainty (k=2)
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage

Range
Uncertainty (k=2)
Range
0 to 120 mV
Uncertainty (k=2)
1,0 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance

Range
Uncertainty (k=2)
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm
Range
Uncertainty (k=2)

Remarks/ref. to internal procedure: BL08.06.03

DC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 mV
Uncertainty (k=2)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (k=2)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (k=2)
17 - 71 µV
Range
11 to 22 V
Uncertainty (k=2)
74 - 140 µV
Range
22 to 220 V
Uncertainty (k=2)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (k=2)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

AC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (k=2)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (k=2)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (k=2)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (k=2)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (k=2)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (k=2)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (k=2)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (k=2)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (k=2)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (k=2)
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,90 - 4,7 µA
Range
0,12 to 1 A
Uncertainty (k=2)
0,023 - 0,12 mA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (k=2)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (k=2)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (k=2)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (k=2)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (k=2)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (k=2)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (k=2)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Calibration of resistance bridge

Range
Uncertainty (k=2)
Range
0,25 to 4 ohm/ohm
Uncertainty (k=2)
2 - 17 µohm/ohm

Remarks/ref. to internal procedure: BL08.06.02

Termocouple (TC) Measure, cal of source

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (k=2)
Range
-200 °C to 630 °C
Uncertainty (k=2)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Source, cal of meter

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01
Permanent facility K03 Time and frequency

Periode, periode time

Range
Uncertainty (k=2)
Range
0,1µs to 100ms
Uncertainty (k=2)
1,0E-9 * Ptime+trigger error

Remarks/ref. to internal procedure: BL21.04.05

Number, number of pulses

Range
Uncertainty (k=2)
Range
1 to 10G
Uncertainty (k=2)
0,60

Remarks/ref. to internal procedure: BL21.04.06

Frequency, as measured quantity

Range
Uncertainty (k=2)
Range
0,5Hz to 5GHz
Uncertainty (k=2)
1,0E-9*f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, as source

Range
Uncertainty (k=2)
Range
1,5 and 10MHz
Uncertainty (k=2)
1,0E-9 * f

Remarks/ref. to internal procedure: BL21.04.03

Time interval

Range
Uncertainty (k=2)
Range
1 to 200 hour
Uncertainty (k=2)
0,80s

Remarks/ref. to internal procedure: BL21.04.04

Frequency, calibration of meter

Range
Uncertainty (k=2)
Range
0,010Hz to 2,0MHz
Uncertainty (k=2)
5,0 µHz + 1,7 ppm

Remarks/ref. to internal procedure: BL08.04.01

Frequency, calibration of source

Range
Uncertainty (k=2)
Range
3Hz to 1MHz
Uncertainty (k=2)
50 µHz + 50 ppm

Remarks/ref. to internal procedure: BL08.05.01
Permanent facility K04 Dimension

Thread gauge, plug, simple pitch diameter

Range
Uncertainty (k=2)
Range
2,2 to 100 mm
Uncertainty (k=2)
3,5 µm
Range
100 to 150 mm
Uncertainty (k=2)
4,0 µm

Remarks/ref. to internal procedure: BL14.05.03

Thread gauge, ring, simple pitch diameter

Range
Uncertainty (k=2)
Range
2,2 to 100 mm
Uncertainty (k=2)
5,0 µm
Range
100 to 150 mm
Uncertainty (k=2)
5,4 µm

Remarks/ref. to internal procedure: BL14.05.04

Caliper

Range
Uncertainty (k=2)
Range
0 to 1000 mm
Uncertainty (k=2)
14 µm + 1,6E-5*L

Remarks/ref. to internal procedure: BL14.04.01 (L = measured length). Depth measurement is not covered by scope of accreditation.

Dial gauge

Range
Uncertainty (k=2)
Range
0 to 1 mm
Uncertainty (k=2)
0,5 µm
At resolution: 0.001 mm
Range
0 to 100 mm
Uncertainty (k=2)
1,2 µm + 1,1E-4*L

Remarks/ref. to internal procedure: BL14.04.04

Gauge block, steel

Range
Uncertainty (k=2)
Range
0,5 to 100 mm
Uncertainty (k=2)
0,13 µm +1,2E-6*L

Remarks/ref. to internal procedure: BL14.07.01

Length, external

Range
Uncertainty (k=2)
Range
0 to 600 mm
Uncertainty (k=2)
0,55 µm + 3,0E-6*L
Range
600 to 1000 mm
Uncertainty (k=2)
2,1 µm + 3,5E-6*L

Remarks/ref. to internal procedure: BL14.05.01

Micrometer, external

Range
Uncertainty (k=2)
Range
0 to 500 mm
Uncertainty (k=2)
1,1 µm + 2,1E-05*L

Remarks/ref. to internal procedure: BL14.04.02
Permanent facility K05 Mechanical quantities

Weight/Mass

Range
Uncertainty (k=2)
Range
0,1 g to 120 g
Uncertainty (k=2)
± 0,26 mg + 1,0 mg/kg
Range
120 g to 2300 g
Uncertainty (k=2)
± 0,60 mg + 2,0 mg/kg
Range
2,3 kg to 20 kg
Uncertainty (k=2)
± 10 mg + 3,0 mg/kg

Remarks/ref. to internal procedure: BL06.07.01

Pressure (abs)

Range
Uncertainty (k=2)
Range
20 to 100 hPa
Uncertainty (k=2)
2,4 Pa + 200 ppm
Range
0,01 to 7,6 Mpa
Uncertainty (k=2)
0,20 Pa + 30 ppm
Range
7,6 to 100 Mpa
Uncertainty (k=2)
75 Pa + 34 ppm
Range
100 to 200 Mpa
Uncertainty (k=2)
75 Pa + 39 ppm
Range
200 to 500 Mpa
Uncertainty (k=2)
500 Pa + 150 ppm

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01

Pressure (diff)

Range
Uncertainty (k=2)
Range
0 to 0,5 Mpa
Uncertainty (k=2)
8,0 - 58 Pa

Remarks/ref. to internal procedure: BL06.05.02/ 06.08.02

Dividing ratio -oil and gas

Range
Uncertainty (k=2)
Range
0,01 to 10 Mpa
Uncertainty (k=2)
50 ppm

Remarks/ref. to internal procedure: BL06.04.04

Kn value -oil and gas

Range
Uncertainty (k=2)
Range
0,01 to 7,6 Mpa
Uncertainty (k=2)
35 ppm

Remarks/ref. to internal procedure: BL06.04.06

Piston/sylinder area -oil and gas

Range
Uncertainty (k=2)
Range
0,01 to 7,6 Mpa
Uncertainty (k=2)
30 ppm
Range
7,6 to 100 Mpa
Uncertainty (k=2)
30 ppm
Range
100 to 200 Mpa
Uncertainty (k=2)
34 ppm

Remarks/ref. to internal procedure: BL06.04.01/06.04.05

Pressure (gauge)

Range
Uncertainty (k=2)
Range
-0,095 to 0 Mpa
Uncertainty (k=2)
20 Pa
Range
0 to 100 hPa
Uncertainty (k=2)
0,3 Pa + 75 ppm
Range
0,01 to 7,6 Mpa
Uncertainty (k=2)
0,20 Pa + 30 ppm
Range
7,6 to 100 Mpa
Uncertainty (k=2)
75 Pa + 34 ppm
Range
100 to 200 Mpa
Uncertainty (k=2)
75 Pa + 39 ppm
Range
200 to 500 Mpa
Uncertainty (k=2)
500 Pa + 150 ppm

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01
Permanent facility K10 Temperature, humidity and thermophysical properties

Resistance termometer

Range
Uncertainty (k=2)
Range
-80 to 300 °C
Uncertainty (k=2)
18 mK +|3 mK/100°C|

Remarks/ref. to internal procedure: BL07.04.02

Indicator

Range
Uncertainty (k=2)
Range
-80 to 300 °C
Uncertainty (k=2)
18 mK +|3 mK/100°C|
Range
300 to 660 °C
Uncertainty (k=2)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.06.01

Transmitter

Range
Uncertainty (k=2)
Range
-80 to 300 °C
Uncertainty (k=2)
18 mK +|3 mK/100°C|
Range
300 to 660 °C
Uncertainty (k=2)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.05.01

Dry block calibrator

Range
Uncertainty (k=2)
Range
-80 to 660°C
Uncertainty (k=2)
50-300 mK

Remarks/ref. to internal procedure: BL07.08.03
The administrative / geographical unit:
Trescal Norway AS avd Tananger field calibration
NorSea Base
Bygg 53
4056 Tananger



Field testing laboratory K01 Electrical quantities

DC power, calibration of meter

Range
Uncertainty (k=2)
Range
1 to 100 mW
Uncertainty (k=2)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (k=2)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (k=2)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (k=2)
Range
0,19 to 1,1 nF
Uncertainty (k=2)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (k=2)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (k=2)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (k=2)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (k=2)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (k=2)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (k=2)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (k=2)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (k=2)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (k=2)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (k=2)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (k=2)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (k=2)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (k=2)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurment

Range
Uncertainty (k=2)
Range
1 to 100 mohm
Uncertainty (k=2)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (k=2)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (k=2)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (k=2)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (k=2)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (k=2)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (k=2)
Range
0 ohm
Uncertainty (k=2)
39 µohm
Range
0 to 1 ohm
Uncertainty (k=2)
85 µohm
Range
1 to 1,9 ohm
Uncertainty (k=2)
160 µohm
Range
1,9 to 10 ohm
Uncertainty (k=2)
250 µohm
Range
10 to 25 ohm
Uncertainty (k=2)
74 µohm
Range
25 to 19 ohm
Uncertainty (k=2)
450 µohm
Range
19 to 100 ohm
Uncertainty (k=2)
180 µohm
Range
100 to 190 ohm
Uncertainty (k=2)
2,9 mohm
Range
190 to 1 kohm
Uncertainty (k=2)
16 mohm
Range
1 to 1,9 kohm
Uncertainty (k=2)
30 mohm
Range
1,9 to 10 kohm
Uncertainty (k=2)
160 mohm
Range
10 to 19 kohm
Uncertainty (k=2)
300 mohm
Range
19 to 100 kohm
Uncertainty (k=2)
1,6 ohm
Range
100 to 1 Mohm
Uncertainty (k=2)
3,0 ohm
Range
1 to 1,9 Mohm
Uncertainty (k=2)
35 ohm
Range
1,9 to 10 Mohm
Uncertainty (k=2)
350 ohm
Range
10 to 19 Mohm
Uncertainty (k=2)
810 ohm
Range
19 to 100 Mohm
Uncertainty (k=2)
10 kohm
Range
0,1 to 1,1 Gohm
Uncertainty (k=2)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 2,2 mV
Uncertainty (k=2)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (k=2)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (k=2)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (k=2)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (k=2)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (k=2)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (k=2)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Voltage, calibration of source

Range
Uncertainty (k=2)
Range
0 to 100 mV
Uncertainty (k=2)
2,7 - 5,6 µV
Range
0,1 to 1 V
Uncertainty (k=2)
7,4 - 25 µV
Range
1 to 10 V
Uncertainty (k=2)
57 - 220 µV
Range
10 to 100 V
Uncertainty (k=2)
0,76 - 3,4 mV
Range
100 to 1000 V
Uncertainty (k=2)
11 - 40 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 mV
Uncertainty (k=2)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (k=2)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (k=2)
17 - 71 µV
Range
11 to 22 V
Uncertainty (k=2)
74 - 140 µV
Range
22 to 220 V
Uncertainty (k=2)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (k=2)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

AC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (k=2)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (k=2)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (k=2)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (k=2)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (k=2)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (k=2)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (k=2)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (k=2)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (k=2)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (k=2)
Range
0 to 10 mA
Uncertainty (k=2)
50 - 250 nA
Range
10 to 30 mA
Uncertainty (k=2)
0,73 - 1,1 µA
Range
30 to 100 mA
Uncertainty (k=2)
15 - 41 µA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (k=2)
Range
0 to 220 µA
Uncertainty (k=2)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (k=2)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (k=2)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (k=2)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (k=2)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (k=2)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (k=2)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (k=2)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Measure, cal of source

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (k=2)
Range
-200 °C to 630 °C
Uncertainty (k=2)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Source, cal of meter

Range
Uncertainty (k=2)
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01
Field testing laboratory K05 Mechanical quantities

Balances

Range
Uncertainty (k=2)
Range
0,1 to 120 g
Uncertainty (k=2)
0,16-0,52 mg
Range
120 to 5000 g
Uncertainty (k=2)
2,0-17 mg
Range
5 to 16 kg
Uncertainty (k=2)
31-64 mg
Range
16 to 60 kg
Uncertainty (k=2)
0,48-1,4 g

Remarks/ref. to internal procedure: BL06.07.02

Pressure (gauge)

Range
Uncertainty (k=2)
Range
-0,095 to 0 Mpa
Uncertainty (k=2)
73 - 40 Pa
Range
0 to 0,35MPa
Uncertainty (k=2)
1,7 - 43 Pa
Range
0,35 to 6MPa
Uncertainty (k=2)
0,72 - 7,5 hPa
Range
6 to 350MPa
Uncertainty (k=2)
0,75 - 53 kPa

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01
Field testing laboratory K10 Temperature, humidity and thermophysical properties

Indicator

Range
Uncertainty (k=2)
Range
-25 to 100 °C
Uncertainty (k=2)
80 mK
Range
100 to 200 °C
Uncertainty (k=2)
100 mK
Range
200 to 660 °C
Uncertainty (k=2)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.06.01

Transmitter

Range
Uncertainty (k=2)
Range
-25 to 100 °C
Uncertainty (k=2)
80 mK
Range
100 to 200 °C
Uncertainty (k=2)
100 mK
Range
200 to 660 °C
Uncertainty (k=2)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.05.01

Calibration of ovens and furnaces (Sensor Accuracy Test)

Range
Uncertainty (k=2)
Range
50 til 650 °C
Uncertainty (k=2)
±0,60 til ±0,90 °C

Remarks/ref. to internal procedure: BL07.08.04

Calibration of ovens and furnaces (Temperature Uniformity Survey)

Range
Uncertainty (k=2)
Range
50 til 650 °C
Uncertainty (k=2)
±0,60 til ±0,90 °C

Remarks/ref. to internal procedure: BL07.08.01