Energy (and power)
Active, Reactive, Apparent
1 and 3 phase
Range
See remark
Uncertainty (k=2)
0,040 %
Remarks/ref. to internal procedure: BL08.07.03 Based on "Statnetts krav til måling av sentralnettsutveksling v.1.1(2016)" and includes evaluation of metering point.
cosφ = -1 til 1
f = 45-70Hz
I = 0,01-120A
U = 5-520V
DC power, calibration of meter
Range
1 to 100 mW
Uncertainty (k=2)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (k=2)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (k=2)
0,064 - 0,16 %
Remarks/ref. to internal procedure: BL08.04.01
Capacitance, calibration of meter
Range
0,19 to 1,1 nF
Uncertainty (k=2)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (k=2)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (k=2)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (k=2)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (k=2)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (k=2)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (k=2)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (k=2)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (k=2)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (k=2)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (k=2)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (k=2)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (k=2)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (k=2)
0,65 - 0,92 %, 0-0.2 Hz
Remarks/ref. to internal procedure: BL08.04.01
Resistance, measurement
Range
1 to 100 mohm
Uncertainty (k=2)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (k=2)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (k=2)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (k=2)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (k=2)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (k=2)
0,024 - 2 kohm
Remarks/ref. to internal procedure: BL08.06.01
Resistance, calibration of meter
Range
0 ohm
Uncertainty (k=2)
39 µohm
Range
0 to 1 ohm
Uncertainty (k=2)
85 µohm
Range
1 to 1,9 ohm
Uncertainty (k=2)
160 µohm
Range
1,9 to 10 ohm
Uncertainty (k=2)
250 µohm
Range
10 to 25ohm
Uncertainty (k=2)
74 µohm
Range
25 to 19ohm
Uncertainty (k=2)
450 µohm
Range
19 to 100ohm
Uncertainty (k=2)
180 µohm
Range
100 to 190ohm
Uncertainty (k=2)
2,9 mohm
Range
190 to 1kohm
Uncertainty (k=2)
16 mohm
Range
1 to 1,9kohm
Uncertainty (k=2)
30 mohm
Range
1,9 to 10kohm
Uncertainty (k=2)
160 mohm
Range
10 to 19kohm
Uncertainty (k=2)
300 mohm
Range
19 to 100kohm
Uncertainty (k=2)
1,6 ohm
Range
100 to 1Mohm
Uncertainty (k=2)
3,0 ohm
Range
1 to 1,9Mohm
Uncertainty (k=2)
35 ohm
Range
1,9 to 10Mohm
Uncertainty (k=2)
350 ohm
Range
10 to 19Mohm
Uncertainty (k=2)
810 ohm
Range
19 to 100Mohm
Uncertainty (k=2)
10 kohm
Range
0,1 to 1,1Gohm
Uncertainty (k=2)
0,4 - 12 %
Remarks/ref. to internal procedure: BL08.04.01
AC Voltage, calibration of meter
Range
0 to 2,2 Mv
Uncertainty (k=2)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (k=2)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (k=2)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (k=2)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (k=2)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (k=2)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (k=2)
19 - 91 mV, 40-01k Hz
Remarks/ref. to internal procedure: BL08.04.01
DC Current
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,9 - 4,7 µA
Range
0,12 to 1,0 A
Uncertainty (k=2)
0,023 - 0,12mA
Remarks/ref. to internal procedure: BL08.05.01
DC Voltage
Range
0 to 120 mV
Uncertainty (k=2)
1,0 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV
Remarks/ref. to internal procedure: BL08.05.01
DC Resistance
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm
Remarks/ref. to internal procedure: BL08.06.03
Energy (and power) Active, Reactive, Apparent 1 and 3 phase
Range
See remark
Uncertainty (k=2)
0,020 %
Remarks/ref. to internal procedure: BL08.07.04
cosφ = -1 til 1
f = 45-70Hz
I = 0,01-120A
U = 5-520V
DC Resistance, calibration of source
Range
0 to 12 ohm
Uncertainty (k=2)
0,05 - 0,23 mohm
BL08.06.03
Range
12 to 120 ohm
Uncertainty (k=2)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (k=2)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (k=2)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (k=2)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (k=2)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (k=2)
160 - 700 ohm
Remarks/ref. to internal procedure: BL08.06.03
DC Voltage, calibration of source
Range
0 to 120 mV
Uncertainty (k=2)
1 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (k=2)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (k=2)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (k=2)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (k=2)
0,82 - 6,1 mV
Remarks/ref. to internal procedure: BL08.05.01
DC Voltage, calibration of meter
Range
0 to 220 mV
Uncertainty (k=2)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (k=2)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (k=2)
17 - 71 µV
Range
11 to 22 V
Uncertainty (k=2)
74 - 140 µV
Range
22 to 220 V
Uncertainty (k=2)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (k=2)
2,3 - 9,8 mV
Remarks/ref. to internal procedure: BL08.04.01
DC Current, calibration of source
Range
0 to 120 µA
Uncertainty (k=2)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (k=2)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (k=2)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (k=2)
0,90 - 4,7 µA
Range
0,12 to 1 A
Uncertainty (k=2)
0,023 - 0,12 mA
Remarks/ref. to internal procedure: BL08.05.01
DC Current, calibration of meter
Range
0 to 220 µA
Uncertainty (k=2)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (k=2)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (k=2)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (k=2)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (k=2)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (k=2)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (k=2)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (k=2)
0,092 - 4,8 A
Remarks/ref. to internal procedure: BL08.04.01
Voltage ratio
Range
HF Powermeter DC verification
Uncertainty (k=2)
Range
0 to 100 % scale
Uncertainty (k=2)
0,06 - 0,6 % point
Zero carryover (0% scale) and instrumentation (100 % scale)
Remarks/ref. to internal procedure: MP 220
AC Current, calibration of meter
Range
0 to 220 µA
Uncertainty (k=2)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (k=2)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (k=2)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (k=2)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (k=2)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (k=2)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (k=2)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (k=2)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (k=2)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (k=2)
2,6 - 4,8 A, 45-100 Hz
Remarks/ref. to internal procedure: BL08.04.01
Energy (and power) Active, Reactive, Apparent
1 and 3 phase
Range
See remark
Uncertainty (k=2)
0,10 %
Remarks/ref. to internal procedure: BL08.07.01
cosφ = -1 til 1
f = 45 Hz til 70Hz
I = 0,01 A til 120 A
U = 5 V til 520 V
Power, using power sensor
Range
0,9 to 1,1 mW
Uncertainty (k=2)
3,1 µW
49 - 51 MHz in 50 ohm system and N connector.
Range
+20 to -70 dBm
Uncertainty (k=2)
0,01 - 0,69 dB
100 kHz to 18 GHz in 50 ohm system and N connector
Range
+20 to -30 dBm
Uncertainty (k=2)
0,01 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector
Remarks/ref. to internal procedure: MP 210 Measured using Power Sensor
Attenuation
Range
0 to 50 dB
Uncertainty (k=2)
0,02 - 0,46 dB
100 kHz to 18 GHz in 50 ohm system with N connector
Remarks/ref. to internal procedure: MP 212 RF substitution
Cal Factor for power sensor with N-connector
Range
80 to 110 %
Uncertainty (k=2)
0,3 - 0,8 % points
100 kHz to 18 GHz in 50 ohm system and N connector
Range
80 to 110 %
Uncertainty (k=2)
Range
Uncertainty (k=2)
0,4 - 1,8 % points
10 MHz to 26,5 GHz in 50 ohm system and 3,5 mm connector
Remarks/ref. to internal procedure: MP 213
Power, generated level using power splitter
Range
1 mW (0 dBm)
Uncertainty (k=2)
3,1 µW
50 MHz in 50 ohm system with N connector
Range
+20 to - 70 dBm
Uncertainty (k=2)
0,04 - 0,68 dB
100 kHz to 18 GHz in 50 ohm system and N connector. The uncertainty will be greater if adaptors must be used.
Range
+10 to - 30 dBm
Uncertainty (k=2)
0,02 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector.
Remarks/ref. to internal procedure: MP 211
Energy (and power) Active,
Reactive, Apparent
1 and 3 phase
Range
See remark
Uncertainty (k=2)
0,10%
Remarks/ref. to internal procedure: BL08.07.01
cosφ = -1 til 1
f = 45 Hz til 70Hz
I = 0,01 A til 120 A
U = 5 V til 520 V
Termocouple (TC) Source, cal of meter
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C
Remarks/ref. to internal procedure: BL08.04.01
Termocouple (TC) Measure, cal of source
Range
-250 °C to 2316 °C
Uncertainty (k=2)
From U=0,11 °C to 0,65 °C
Remarks/ref. to internal procedure: BL08.05.01
Resistance thermometer Source, cal of meter
Range
-200 °C to 630 °C
Uncertainty (k=2)
From U=0,02 °C to 0,18 °C
Remarks/ref. to internal procedure: BL08.04.01